| Sign In | Join Free | My benadorassociates.com |
|
| Categories | Sleep Apnea Monitor |
|---|---|
| Place of Origin: | China |
| Brand Name: | BERRY |
| Model Number: | BM2000D |
| Packaging Details: | Size:35(H) x 35(W) x 20(D)mm Weight:20g |
| MOQ: | 2 units |
| Price: | $80.00/units 2-49 units |
| Power Source: | Electricity |
| Warranty: | 1 Year |
| After-sale Service: | Online technical support |
| Power Supply Mode: | Built-in Battery |
| Material: | Plastic |
| Shelf Life: | 1years |
| Quality Certification: | ce |
| Instrument classification: | Class II |
| Safety standard: | None |
| Type: | Blood Testing Equipments |
| Display: | OLED |
| SPO2: | 35%-100% |
| SPO2 Accuracy: | ±2%(80%-100%);±3%(70%-79%) |
| Pulse Rate: | 25-250bpm |
| Working voltage: | D.C.3.4V-D.C.4.3V |
| Battery Type: | one 3.7V lithium battery |
| Weight: | 80g |
| Certificate: | ISO13485 |
| Company Info. |
| Shanghai Berry Electronic Tech Co., Ltd. |
| Verified Supplier |
| View Contact Details |
| Product List |


Display SpO2, PR | Battery Type: 3.7v (600mAh) Lithium battery |
Pulse Rate: Measurement range: 25 - 250bpm Accuracy: ±2bpm | SpO2: Measurement range: 0% - 100% Accuracy: ±2% (70% - 100%) |
Size:55(L)x37(W)x18(D)mm | Weight: 52g±1g (Do not contain sensor) |
Working Voltage: DC 3.4v ~ DC 4.3v | Free App: “BerrySmartHealth” |


Comparison: Reference Pulse Oximeter | ||||||||||
Berry Pulse Oximeter | 70 ~100 | 90 ~ 100 | 80 - < 90 | 67 - < 80 | RESE≤3% Within the range of 70~100% | |||||
# pts | 240 | 79 | 81 | 81 | ||||||
Error | -0.6 | -1.1 | -0.9 | 0.1 | Pass | |||||
Root mean square error | 1.9 | 1.7 | 2.0 | 2.1 | ||||||





Mean Blood Oxygen | ||||
Average blood oxygen calculation accuracy: 98.51% | ||||
Diagnosis of hypoxemia | ||||
Sensitivity: 89.29% | Specificity: 77.78% | Accuracy:92.59% | ||
OSAHS Judgment Results | ||||
Sensitivity: 94.17% | Specificity: 61.21% | Accuracy:89.62% | ||






Display SpO2, PR | Battery Type: 3.7v (600mAh) Lithium battery |
Pulse Rate: Measurement range: 25 - 250bpm Accuracy: ±2bpm | SpO2: Measurement range: 0% - 100% Accuracy: ±2% (70% - 100%) |
Size:55(L)x37(W)x18(D)mm | Weight: 52g±1g (Do not contain sensor) |
Working Voltage: DC 3.4v ~ DC 4.3v | Free App: “BerrySmartHealth” |


Comparison: Reference Pulse Oximeter | ||||||||||
Berry Pulse Oximeter | 70 ~100 | 90 ~ 100 | 80 - < 90 | 67 - < 80 | RESE≤3% Within the range of 70~100% | |||||
# pts | 240 | 79 | 81 | 81 | ||||||
Error | -0.6 | -1.1 | -0.9 | 0.1 | Pass | |||||
Root mean square error | 1.9 | 1.7 | 2.0 | 2.1 | ||||||





Mean Blood Oxygen | ||||
Average blood oxygen calculation accuracy: 98.51% | ||||
Diagnosis of hypoxemia | ||||
Sensitivity: 89.29% | Specificity: 77.78% | Accuracy:92.59% | ||
OSAHS Judgment Results | ||||
Sensitivity: 94.17% | Specificity: 61.21% | Accuracy:89.62% | ||




|