| Sign In | Join Free | My benadorassociates.com |
|
| Categories | ETCO2 Sensor |
|---|---|
| Place of Origin: | China |
| Brand Name: | Med Accessories |
| Model Number: | MCMZ30 |
| Packaging Details: | PE+box or customized |
| Certification: | CE,ISO,SFDA |
| Delivery Time: | About one week |
| Payment Terms: | T/T |
| MOQ: | 1pcs |
| Disinfecting Type: | no |
| Size: | 3m |
| Material: | TPU |
| Quality Certification: | CE |
| Safety standard: | ISO13485 |
| OEM Part #: | M2501A |
| Warranty: | 1 year |
| Latex-free: | Yes |
| Patient Size: | Adult/Pediatric |
| Company Info. |
| Med Accessories Technology Dongguan Co., Ltd. |
| Verified Supplier |
| View Contact Details |
| Product List |
M2501A Mainstream Capnography HP EtCO2 Sensor for MP5 MP2 X2
| Manufacturer | Model |
|---|---|
| HP Compatible | M3002A MMS X2, M8102A MP2, M8105A MP5 |
| Project | Specifications | |||||||||||||
| Principle of Operation | Non-dispersive infrared(NDIR) single beam optics, dual wavelength, no moving parts. | |||||||||||||
| Initialization Time | Capnogram displayed in less than 3 seconds,full specifications within 3 minutes(At 25°C). | |||||||||||||
| CO2 Measurement Range | 0 - 114 mmHg; 0 - 15 %; 0 - 15.2 kPa. | |||||||||||||
| CO2 Rise Time | Rise Time Less than 80 ms - Adult Reusable or Single-Patient-Use
Airway Adapter Less than 80 ms - Infant Reusable or Single-Patient-Use Airway Adapter. | |||||||||||||
| CO2 Resolution | 0. 1mmHg ; 0 to 38 mmHg.; 0.25 mmHg ; 39 to 114mmHg. | |||||||||||||
| CO2 Accuracy | 0 - 40mmHg ±2 mmHg.; 41 - 76mmHg ±5% of reading.; 77 - 114 mmHg ±8%
of reading. Above 80 breaths per minute ±12% of reading. * Note: Gas emperature at 25°C. | |||||||||||||
| CO2 Stability | Short term drift: Drift over four hours shall not exceed 1 mmHg
maximum. Long term drift: Accuracy specification will be maintained over a 120-hour period. | |||||||||||||
| Sampling Frequency | 100Hz. | |||||||||||||
| Respiration Rate Range | 3 to 150 breaths per minute. | |||||||||||||
| Respiration Rate Accuracy | ±1 breath. | |||||||||||||
| Calibration | No routine user calibration required. An airway adapter zero is
required when changing to a different style of airway adapter. | |||||||||||||
| Airway Adapters | Single-patient-use or reusable, < 5 cc dead space (adult), < 1 cc dead space infant. | |||||||||||||
| Temperature | Operating: 0 to 40°C; Storage: -40 to 70°C; Transport: -40 to 70°C. | |||||||||||||
| Humidity | Operating: 10 to 90% RH, non-condensing; Storage: 10 to 90% RH,
non-condensing.; Transport: 10 to 90% RH, non-condensing. | |||||||||||||
| Atmospheric Pressure | Operating: 400-800 mmHg; Storage: 400-800 mmHg; Transport: 400-800 mmHg. | |||||||||||||
| Water Resistance | IPX4 - Splash-proof - Sensor only. | |||||||||||||
| Voltage Requirements | 5VDC ± 5 %. | |||||||||||||
| Power Rating | <1.0W Typical; <1.8 W Warm-up. | |||||||||||||
| Sensor Weight | <85g. | |||||||||||||
| Sensor Size | Deep*Wide*High:48mm * 24.3mm * 35.5mm. | |||||||||||||
| Cable Length | 2.4m. | |||||||||||||
|