| Sign In | Join Free | My benadorassociates.com |
|
| Categories | Oxygen Gas Sensor |
|---|---|
| Model Number: | OOM202 |
| Place of Origin: | USA |
| MOQ: | 1 |
| Price: | Negotiable |
| Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
| Supply Ability: | 1000pcs/Month |
| Delivery Time: | 3-5workingdays |
| Packaging Details: | Standard Package |
| Type: | Oxygen sensor |
| Use/application: | Industrial/medical |
| Measuring range: | 0%-100% oxygen |
| Pressure range: | 0.6 bar-2 bar |
| Operating temperature: | 0 ° C+50 ° C |
| Response time: | Less than 12 seconds |
| Company Info. |
| ShenzhenYijiajie Electronic Co., Ltd. |
| Verified Supplier |
| View Contact Details |
| Product List |
OOM202 Gas Oxygen Sensors Industrial Medical Excellent Signal Stability
Use the advantages:
• Compliant with European MDD (CE certification)
• Meets ISO 80601-2-55
• Designed and manufactured according to EN ISO 13485
• Accurate and reliable fast response
• Resistant to N2O
• Excellent signal stability
• High product quality
• Short delivery times
• Technical support
• Made in Germany
• FDA cleared
Datasheet
| Measurement range | 0 %-100 % oxygen (at atmospheric pressure) |
| Nominal sensor lifetime | ≥ 1 000 000 % volume oxygen hours |
| Output in ambient air | 13 mV ... 16 mV |
| Electrical interface | 3 pin (Molex® 22-11-1031) |
| Accuracy meets | 80601-2-55 requirements |
| Repeatability | < 1 % volume O2 at constant temperature and pressure |
| Linearity error | < 3 % relative |
| Response time | < 12 s to 90 % of final value |
| Zero offset voltage | < 200 µV in 100 % nitrogen, applied for 5 min |
| Cross interference meets | 80601-2-55 requirements |
| Influence of humidity | -0.03 % rel. per % RH at 25 °C |
| Pressure range | 0.6 bar-2 bar (ppO2 0-1250 mbar O2) |
| Influence of pressure | Proportional to change in oxygen partial pressure |

|