| Sign In | Join Free | My benadorassociates.com |
|
| Categories | Thermal Analyzer |
|---|---|
| Brand Name: | lonroy |
| Place of Origin: | Guangdong, China |
| MOQ: | 1 |
| Thermal conductivity range: | 0.0001 ~ 1W/mk |
| Measurement accuracy: | 1% |
| Sample size: | 50×50 ~ 300×300 (mm) |
| Measurement time: | 15-20min |
| Hot surface temperature: | Room temperature ~ 99.9℃ (standard), up to 499.9℃ |
| Cold surface temperature: | 0 ~ 99.9℃ (standard), -10 ~ 99.9℃ |
| Temperature control accuracy: | 0.1℃ |
| Pressure range: | 0 ~ 1000N |
| Pressure resolution: | 0.1N |
| Sample thickness range: | 0 ~ 50mm |
| Thickness resolution: | 1um |
| Host dimensions: | 550×450×1500mm |
| Host weight: | 100Kg |
| Voltage: | 220 V |
| Protection class: | IP57 |
| Company Info. |
| DONGGUAN LONROY EQUIPMENT CO LTD |
| Verified Supplier |
| View Contact Details |
| Product List |
ASTM C518 High Precision Thermal Insulation Material Conductivity Tester DRPL-IV Thermal Resistance Detecting Instrument
1.use
Using the international popular heat flow meter to detect thermal
conductivity and thermal resistance method, with computer to
achieve automatic detection, generate experimental report. The
instrument uses a stable hot surface temperature added to the
sample side, the heat is transferred to the cold surface through
the sample, and the heat transfer is measured to calculate the
thermal conductivity and thermal resistance. Using imported sensor
and computer technology, good measurement accuracy, high
reliability, good repeatability characteristics. Especially
suitable for low thermal conductivity measurements (accurate to
0.1mw). Can be used for plastics, rubber, insulation materials,
aerogel plate, vacuum glass and other testing, widely used in
colleges and universities, scientific research institutes, quality
inspection, factories and mines.
Reference standard: GB/T10295-2008 (Determination of steady-state
thermal resistance and related properties of insulating materials
Heat flow meter method) and ASTM C518-04 test method for determination of steady-state heat flux and
heat transfer characteristics by heat flow meter method.
ASTM C518 High Precision Thermal Insulation Material Conductivity Tester DRPL-IV Thermal Resistance Detecting Instrument
2.Product Parameter
| thermal conductivity range and accuracy | 0.0001 ~ 1W/mk, can be measured accurately to 0.1mw/mk. |
| measurement accuracy | 1% |
| sample size | 50×50 ~ 300×300 (mm) |
| Measurement time | 15-20min |
1 can test solid materials, powder materials.
2, hot surface temperature and control: room temperature ~ 99.9℃
(standard configuration), room temperature ~ 499.9℃, higher
temperature can be determined according to user requirements. Plate
heater, bidirectional SCR control, temperature control accuracy
0.1℃.
3, cold surface temperature and control: 0 ~ 99.9℃ (standard
configuration),-10 ~ 99.9℃, according to user requirements to
determine the lower temperature. The cold surface adopts water
cooling to ensure the temperature stability of the cold surface
(precision constant temperature sink, temperature adjustable),
temperature control accuracy of 0.1℃.
4, the sample adopts computer precision pressure clamping, pressure
range 0 ~ 1000N, resolution 0.1N.
5, sample thickness automatic measurement, 0 ~ 50mm, resolution
1um.
6,imported heat flow meter, stable and reliable performance.
7. The whole process adopts the computer to automatically test and
print and store experimental data.
ASTM C518 High Precision Thermal Insulation Material Conductivity Tester DRPL-IV Thermal Resistance Detecting Instrument
3.complete set
1. One host computer;
2. A set of analysis software (Chinese and English versions);
3, HX-1005 precision low temperature constant temperature tank one
(temperature control accuracy 0.01℃);
4, a Lenovo computer (users can choose);
5. A standard sample;
4. Dimensions and weight
1. Host: standard landing model: (length × width × height)
550×450×1500mm, net weight: 100Kg.
2, precision low temperature constant temperature tank: (length ×
width × height) 300×350×530mm, net weight: 35Kg.
ASTM C518 High Precision Thermal Insulation Material Conductivity Tester DRPL-IV Thermal Resistance Detecting Instrument
4.Technical services
1, one year free warranty, long-term technical services, software
lifetime free upgrade.
2. Free on-site installation and debugging throughout the country.
All over the country has the company's mobile technical personnel,
can provide technical services.
|