GE Logiq S8 Medical Ultrasound Repair
|
GE Logiq S8 System Reports Error Failure phenomenon 2D image and blood flow of the machine is abnormal, the blood flow knob is not working good. Failure analysis According to the above failure phenomenon, it can be judged that the image channel of the machine is abnormal. Therefore, it is suspected that there is a problem with the BF channel board. It was also found that the knob& of blood flow section was not working good, so the control panel...
Guangzhou Rongtao Medical Tech LTD.
|
GE LOGIQ P8 Voluson S6 S8 LOGIQ S8 Ultrasound Machine Trackball FLT20UWB-G6
|
FLT20UWB-G6 GE LOGIQ P8 Voluson S6 S8 LOGIQ S8 Ultrasound Machine trackball Control Panel Map 1. Power button 2. USB Port x 2 (USB3.0) 3. Touch Panel 4. TGC 5. Reverse key 6. Rotary and Joystick controls 7. User Define keys 8. BT Keys 9. F1: Access Online ......