Categories Stationary Meter Test System
Brand Name: GENY
Model Number: YC-1893R
Certification: ISO, CE
Place of Origin: China
MOQ: 1
Price: negotiate rates
Payment Terms: T/T, L/C, D/P, T/T
Supply Ability: 5 sets / 1 months
Delivery Time: 6~8weeks
Packaging Details: Plywood cases
Structure: All-in-one
Usage: Energy Meter Calibration, Lab Research
Ambient Temperature: -10℃ to +40℃
Power Supply: 3P4W 3×380V±10%, 50Hz±2Hz
Ambient Humidity: 35% to 85% R.H.
Voltage Resolution: 0.001%
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

Customized High Precision Power Supply & Electrical System Manufacturer

Customized High Precision Power Supply & Electrical System Manufacturer

Power Supply / Electrical System


Voltage240/415 V AC supply (+5% & -10%) or customized
Frequency50/60Hz ±2Hz
Power capacity> 3 × 6kVA
Ambient temperature-10°C … +40°C
Relative humidity35% … 95%
System GroundingNeutral solidly grounded
Surge protection level≥ 10 kV/120 Joules (10/1000µs)

Power source


Voltage
Test voltage output (Phase-Neutral)3× (24V – 500V) or customized
Power of the voltage output3×100VA
ResolutionBetter than 0.01% of full scale value of range
Setting accuracyBetter than 0.05% at the final range value
StabilityBetter than 0.005%/h (integration time 150 s)
Load regulation from 0-max. load< 0.01%
Distortion factor< 0.3% for linear resistance load
Harmonics2nd – 21st free programmable
Current
Test current output3× (1mA – 120A)
Power of the current output3×150VA
ResolutionBetter than 0.01% of full scale value of range
Setting accuracyBetter than 0.05% at the final range value
StabilityBetter than 0.005%/h (integration time 150 s)
Load regulation from 0-max. load< 0.01%
Distortion factor< 0.3% for linear resistance load
Harmonics2nd – 21st free programmable
Phase angle
Range0 -- 360°
Resolution0.01°
Setting accuracy0.1°
Frequency
Range45Hz – 65Hz
Resolution0.01Hz

Reference Standard


Power supply
VoltageAC220V ±10%
Frequency50Hz ±2Hz
Max power consumption45VA
Using condition
Environment temperature10ºC … 30ºC
Relative humidity45% … 75%
Outside magnetic intensity≤ 0.025mT
Frequency range45Hz … 65Hz
Test current
Range1mA … 120A
Error

- ±0.02% (50mA…120A) (relative measurement value)

- ±0.05% (1…50mA) (relative range full-scale value)

Current display range1.000mA … 120.000A (1A, 10A, 100A)
Test voltage
Range50V … 480V (auto range)
Error

- ±0.02% (50V…480V) (relative measurement value)

- ±0.05% (5V…50V) (relative measurement value)

Voltage display range50.000(V) … 480.000(V)
Power measure error
Active power+0.02% (relative measurement value) 50V … 480V, 0.05A … 120A
Reactive power+0.05% (relative measurement value) 50V … 480V, 0.05A … 120A
Apparent power+0.05% (relative measurement value) 50V … 480V, 0.05A … 120A
Test Power Factor(PF=P/F) Error: 0.05%
Display range-1.00000 … +1.00000
Energy constant

CH=109 imp/kWh (1A & 0.1A step)

CH=108 imp/kWh (10A step)

CH=107 imp/kWh (100A step)

Measuring modes

- 4 wire active/ reactive/ cross-connect reactive

- 3 wire active/ reactive/ cross-connect reactive

- 2 wire active/ reactive


Energy pulse input
Frequency0.5kHz
Voltage24V
Measurement range of input
OutputTTL pulse, load > 2kΏ
Dimensions (W * H * D)131mm * 432mm * 470mm
Weight16kgs

Regular Features


  • Power Measurement: Measures active/reactive power and energy.
  • Current Testing: Includes start-up and inrush current tests.
  • Efficient Testing: Automatic color marking for quick procedures.
  • Error Analysis: Covers basic, average error, standard deviation, and 24-hour variation.
  • Impact Assessment: Evaluates voltage, frequency, harmonics (2nd to 21st order), and phase sequence effects.
  • Simultaneous Measurement: Handles multiple meters with different constants.
  • Protection: Shields against open-circuit current and short-circuit voltage.
  • Operation Modes: Automatic and manual options.
  • Data Management: Automates error data processing, storage, and output.
  • Custom Reporting: Offers advanced data management and customizable report formats.
  • Barcode Input: Streamlines operations with barcode support.
  • Power Supply: Treats power supply to eliminate beat frequency issues.
  • Customization: Tailors appearance and structure to site specifications.

Special Features


  • Bidirectional Testing: Tests bidirectional active/reactive errors.
  • Precision Time Base: SW-5D source for daily timing error measurement.
  • Communication: Supports testing and broadcast synchronization.
  • Constant Verification: Uses pulse count, time, and start/end energy methods.
  • Time-of-Use: Tests switching, demand period, and demand value errors.
  • Meter Reading: Evaluates meter reading and combination errors.
  • Four-Quadrant Power: Performs four-quadrant reactive power testing.